Application of Remote Dielectric Sensing in Volumetric Evaluation for Maintenance Hemodialysis Patients
Authors' conclusion
ReDS has potential in fluid evaluation for maintenance hemodialysis patients and can serve as a useful tool to assist in clinical hydration status assessment.
Study context
- Population
- Maintenance HD patients
- Participants
- 115 patients
- Outcome
- Correlation of ReDS values with overhydration (bioelectrical impedance analysis) and biochemistry indicators such as brain natriuretic peptide
Main finding
As reported in the abstract
- Pre-dialysis remote dielectric sensing (ReDS) reading, mean ± SD
- 27.47% ± 4.88% (range 16.00% to 39.00%)
- Post-dialysis ReDS value, median
- 24% (range 16% to 37%)
- ReDS distinguishes fluid overload patients (pre-dialysis overhydration value ≥2.5 L)
- p < 0.001
- Correlation of ReDS with overhydration value, pre-dialysis
- R = 0.38, p < 0.001
- Correlation of ReDS with left ventricular ejection fraction
- R = −0.333, p = 0.018
- ROC cut-off of pre-dialysis ReDS value to discriminate pre-dialysis overhydration value ≥2.5 L
- 26.5%
Original publication
Application of Remote Dielectric Sensing in Volumetric Evaluation for Maintenance Hemodialysis Patients. Blood Purif. Published 18 June 2025. doi:10. 1159/ 000546872