Acid–base implications of the Gibbs-Donnan effect during continuous veno-venous hemofiltration
Authors' conclusion
These in-vitro and in-vivo studies demonstrate that albumin linearly affects the sieving coefficient of ions, increasing the strong ion difference of plasma water during its passage through the filter and thus having a systemic alkalizing effect.
Study context
- Population
- Oligo-anuric critically ill patients undergoing continuous veno-venous hemofiltration (CVVH); in-vitro CVVH experiments with and without albumin
- Participants
- 22 plus in-vitro experiments
- Outcome
- Gibbs-Donnan effect across the filter during CVVH and its acid-base implications
Methodological detailsin-vitro and in-vivo, prospective, single-center, PrismaMax machine (Baxter), physical-chemical approach
- In-vitro and in-vivo
- Prospective
- Single-center
- PrismaMax machine (Baxter)
- Physical-chemical approach
Main finding
As reported in the abstract
- In vitro without albumin: sieving coefficient for both positive and negative ions
- 1
- In vitro with albumin: sieving coefficient for sodium changed linearly with albumin concentration
- r = −0.94, p < 0.001
- In vitro with albumin: sieving coefficient for chloride changed linearly with albumin concentration
- r = 0.88, p < 0.001
- In vitro: progressive linear increase in post-filter strong ion difference
- β = 1.1, r = 0.77, p = 0.003
- In vivo: calculated albumin concentration increased across the filter
- 2.2 ± 0.5 g/dL vs. 3.1 ± 0.7 g/dL
- In vivo: sodium retention across the filter
- p < 0.001
- In vivo: chloride excretion across the filter
- p < 0.001
- In vivo: higher post-filter strong ion difference
- p < 0.001
Original publication
Acid–base implications of the Gibbs-Donnan effect during continuous veno-venous hemofiltration. J Nephrol. Published 10 March 2025. doi:10. 1007/ s40620-025-02238-0
Composed by AI from the published abstract and checked against it. How Briefs are made →