Accuracy of validated blood pressure measuring devices for out-of-office measurement in children and adolescents: a systematic review and meta-analysis
Authors' conclusion
Few BP devices are validated for out-of-office use in the pediatric population, with ABPM devices showing higher accuracy than HBPM devices.
Study context
- Population
- Children and adolescents; successfully validated out-of-office blood pressure (BP) measuring devices17 validation studies
- Comparison
- Home blood pressure measurement (HBPM) vs ambulatory blood pressure measurement (ABPM) validated devices
- Outcome
- Accuracy of validated devices per AAMI/ESH/ISO 2018 validation criterion 1 (mean difference for systolic blood pressure (SBP), mmHg)
Methodological detailsAAMI/ESH/ISO Universal Standard (ISO 81060-2:2018), Searched PubMed, Scopus and CENTRAL
- AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)
- Searched PubMed, Scopus and CENTRAL
Pooled result
Absolute mean difference for SBP (criterion 1)
Key notes
- Significantly lower accuracy with HBPM
Further pooled results
mmHg
| Outcome | HBPM | ABPM | P value |
|---|---|---|---|
| Signed mean difference for SBPNo significant difference | 1.34 (95% CI −0.55 to 3.24) | −0.49 (95% CI −1.40 to 0.41) | P = 0.086 |
Heterogeneity
- High heterogeneity in sensitivity and subgroup analyses (I2)
- > 90%
Study quality
- Inadequate reporting on the validation procedure
Original publication
Accuracy of validated blood pressure measuring devices for out-of-office measurement in children and adolescents: a systematic review and meta-analysis. Pediatr Nephrol. Published 27 November 2025. doi:10. 1007/ s00467-025-06974-0