Kidney & hypertension

Accuracy of validated blood pressure measuring devices for out-of-office measurement in children and adolescents: a systematic review and meta-analysis

Pediatr NephrolPublished Systematic review and meta-analysis of validation studies

Authors' conclusion

Few BP devices are validated for out-of-office use in the pediatric population, with ABPM devices showing higher accuracy than HBPM devices.

Study context

Population
Children and adolescents; successfully validated out-of-office blood pressure (BP) measuring devices17 validation studies
Comparison
Home blood pressure measurement (HBPM) vs ambulatory blood pressure measurement (ABPM) validated devices
Outcome
Accuracy of validated devices per AAMI/ESH/ISO 2018 validation criterion 1 (mean difference for systolic blood pressure (SBP), mmHg)
Methodological detailsAAMI/ESH/ISO Universal Standard (ISO 81060-2:2018), Searched PubMed, Scopus and CENTRAL
  • AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)
  • Searched PubMed, Scopus and CENTRAL

Pooled result

Absolute mean difference for SBP (criterion 1)

Absolute MD

1.35 mmHg

95% CI 0.21–2.50

P = 0.023

As reported in the abstract

Further pooled results

mmHg

OutcomeHBPMABPMP value
Signed mean difference for SBPNo significant difference1.34 (95% CI −0.55 to 3.24)−0.49 (95% CI −1.40 to 0.41)P = 0.086

Heterogeneity

High heterogeneity in sensitivity and subgroup analyses (I2)
> 90%

Study quality

Inadequate reporting on the validation procedure

Original publication

Accuracy of validated blood pressure measuring devices for out-of-office measurement in children and adolescents: a systematic review and meta-analysis. Pediatr Nephrol. Published 27 November 2025. doi:10.1007/s00467-025-06974-0