Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis

Perit Dial IntPublished Diagnostic accuracy study

Authors' conclusion

We demonstrated satisfactory performance of the FSL monitoring system by both analytical and clinical metrics in this cohort of PD patients using treatment prescriptions including Icodextrin-based fluids.

Study context

Population
People with diabetes on peritoneal dialysis (PD) using Icodextrin in their PD regime
Participants
12 people
Comparison
FreeStyle Libre (FSL) vs venous glucose on Yellow Springs Instrument (YSI) and vs SMBG
Outcome
Analytical and clinical accuracy of the FSL continuous glucose monitoring system
Follow-up
10 days
Methodological detailssingle-centre, observational, blinded research model of the FSL
  • Single-centre
  • Observational
  • Blinded research model of the FSL

Performance

As reported in the abstract

Mean absolute relative difference of FSL
Test or modelValueMatched pairs
Yellow Springs InstrumentYSImean 9.8% (95% CI 8.6–11.1)84
SMBGmean 17.3% (95% CI 16.24–18.43)416

Value

Yellow Springs InstrumentYSImean 9.8% (95% CI 8.6–11.1)
SMBGmean 17.3% (95% CI 16.24–18.43)

Matched pairs

Yellow Springs InstrumentYSI84
SMBG416

Further findings

Clinical accuracy: sensor values in Parkes consensus error grid zones A and B

OutcomeYSISMBG
Systematic error of FSL (Bland-Altman)−0.6 ± 1.0 mmol/l−1.4 ± 1.9 mmol/l

Original publication

Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis. Perit Dial Int. Published 23 December 2025. doi:10.1177/08968608251406514

Composed by AI from the published abstract and checked against it. How Briefs are made →