Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis
Authors' conclusion
We demonstrated satisfactory performance of the FSL monitoring system by both analytical and clinical metrics in this cohort of PD patients using treatment prescriptions including Icodextrin-based fluids.
Study context
- Population
- People with diabetes on peritoneal dialysis (PD) using Icodextrin in their PD regime
- Participants
- 12 people
- Comparison
- FreeStyle Libre (FSL) vs venous glucose on Yellow Springs Instrument (YSI) and vs SMBG
- Outcome
- Analytical and clinical accuracy of the FSL continuous glucose monitoring system
- Follow-up
- 10 days
Methodological detailssingle-centre, observational, blinded research model of the FSL
- Single-centre
- Observational
- Blinded research model of the FSL
Performance
As reported in the abstract
Mean absolute relative difference of FSL
| Test or model | Value | Matched pairs |
|---|---|---|
| Yellow Springs InstrumentYSI | mean 9.8% (95% CI 8.6–11.1) | 84 |
| SMBG | mean 17.3% (95% CI 16.24–18.43) | 416 |
Value
Yellow Springs InstrumentYSImean 9.8% (95% CI 8.6–11.1)
SMBGmean 17.3% (95% CI 16.24–18.43)
Matched pairs
Yellow Springs InstrumentYSI84
SMBG416
Further findings
Clinical accuracy: sensor values in Parkes consensus error grid zones A and B
| Outcome | YSI | SMBG |
|---|---|---|
| Systematic error of FSL (Bland-Altman) | −0.6 ± 1.0 mmol/l | −1.4 ± 1.9 mmol/l |
Original publication
Accuracy of a continuous glucose monitoring system in people with diabetes on peritoneal dialysis. Perit Dial Int. Published 23 December 2025. doi:10. 1177/ 08968608251406514