A multi-institutional, observational study of outcomes after catheter placement for peritoneal dialysis in Japan
An erratum to this article was published (Perit Dial Int, Mar 2024 ↗)
Authors' conclusion
PD catheter placement in Japan was proven to be safe and appropriate.
Study context
- Population
- Adult patients who underwent first peritoneal dialysis (PD) catheter placement for ESKD in Japan between April 2019 and March 2021401 with externalised catheters
- Participants
- 409 patients
- Outcome
- Time to catheter patency failure and technique failure, and perioperative infectious complications within 30 days of catheter placement
Methodological detailsCox proportional hazards model, logistic regression
- Cox proportional hazards model
- Logistic regression
Main finding
As reported in the abstract
Catheter patency failure, technique failure and perioperative infectious complications within 30 days · 12 months
Estimate (95% CI)
Serum albuminCox proportional hazards model; independent risk factor
HR for technique failure 0.44(0.27–0.70)
Straight type catheterCox proportional hazards model; independent risk factor
HR for technique failure 2.14(1.24–3.69)
Diabetes mellituslogistic regression; only independent risk factor
OR for perioperative infectious complications 2.70(1.30–5.58)
- Catheter patency failure
- 25 (6.2%) of 401
- Technique failure at 12 months (cumulative incidence function)
- 15.3%
Safety
Perioperative complications
- Occurrence rate generally met the audit criteria of the International Society for Peritoneal Dialysis (ISPD) guidelines
Original publication
A multi-institutional, observational study of outcomes after catheter placement for peritoneal dialysis in Japan. Perit Dial Int. Published 26 August 2023. doi:10. 1177/ 08968608231193240
Composed by AI from the published abstract and checked against it. How Briefs are made →